2-38
SYSTEM MEMORY OVERVIEW
SYSY
SYSY
SY
STEM MEMORSTEM MEMOR
STEM MEMORSTEM MEMOR
STEM MEMOR
YY
YY
Y
SYSTEM MEMORY OVERVIEW
Various types of information are saved in the system, using
different types of memory.
The program that runs the unit is stored in Flash memory.
This allows it to be field upgraded via the RS232 Port.
The Datalog, User, Administrator and Factory parameters are
saved in EEPROM memory. This is a permanent memory
(10 years minimum life) that maintains the values, even when
the power is off.
These saves occur automatically at various points in the
program and there is a short wait (1-2 seconds) while this
save is taking place. When active, the Save Mode screen will
appear.
**SAVE MODE**
The Model information and Datalog data is stored in battery
backed up RAM. This is a permanent memory (10 year
minimum life) that maintains the data, even with the power off.
The saving of this data is very fast and completely under
program control so no display indications are given.