BC Biomedical IPA-2000S Manual do Utilizador Página 111

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 162
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 110
4-1
PART 4 : INFUSION ANALYZER--DISTAL OCCLUSION
INFUSION INFUSION
INFUSION INFUSION
INFUSION
ANAN
ANAN
AN
ALAL
ALAL
AL
YZER--OCCLYZER--OCCL
YZER--OCCLYZER--OCCL
YZER--OCCL
USIONUSION
USIONUSION
USION
PART 4 : INFUSION ANALYZER--DISTAL OCCLUSION
There are two basic types of occlusion tests that may be run:
Distal and Proximal.
The Distal test is performed by the system by closing an
internal solenoid to occlude the flow and then monitoring and
recording the pressure versus time. The test is terminated
either manually or, if the Nurse Call connection is made,
automatically. If used, the time of the Nurse Call is noted and
three additional pressure readings are taken to insure the
pump turned off.
The Proximal test must be performed manually since the unit
has no control over the upstream fluid path. This option is
provided in the unit as a convenient timer for the test. No
data is recorded in the Datalog during this test.
Included are:
CAn overview and step by step instructions for
running a Distal Occlusion test
CAn overview and step by step instructions for
running a Proximal Occlusion test
CDetails for using the Datalog
Vista de página 110
1 2 ... 106 107 108 109 110 111 112 113 114 115 116 ... 161 162

Comentários a estes Manuais

Sem comentários