ESU Analyzer Series
ESU Analyzer Series
S
S
PECIFICATIONS
PECIFICATIONS
Flip for product pricing - Page F
Flip for product pricing - Page F
Optional: External Load
Optional: External Load
Precision Power Resistor
Precision Power Resistor
(See Page F for a
(See Page F for a
listing of available values)
listing of available values)
The ESU-2300 allows for the use of
External or a combination of Internal &
External Load Resistors to cover ranges
outside the provided Internal Load Range.
Setup Parameters allow the user to tell the
unit about External Loads. This ensures
the continued accurate measurement &
display of power.
INTERNAL LOAD VALUES NOT
EXACTLY WHAT YOU NEED?
Measurement
Method
Industry Standard Current
Sensing, using RF Current
Transformer (Pearson Coil)
Power
Range 1.0 to 400.0 Watts RMS
Resolution 0.1 Watts
Accuracy
5% Reading or ± 3 Watts
(whichever is greater)
Current
Range 20 to 2500 mA RMS
Resolution 1mA
Accuracy 2.5% Reading or ± 15 mA
Limits
Bandwidth 10 kHz to 10 MHz
Crest Factor 1.4 to 500
Voltage 10,000 V Peak
Main Test Load
Range 50 to 750 Ohms
Resolution 50 Ohms
Accuracy 1% (DC)
Duty Cycle (10 SEC ON, 30 SEC OFF)
Auxiliary Test Load
Fixed 200 Ohms
Accuracy 1% (DC)
CQM Test Load
Range 1 to 500 Ohms
Resolution 1 Ohms
Accuracy 2% or ± 2 Ohms
Physical
Enclosure
High Impact Plastic, UL 94 V-0
Face-Lexan, Back Printed
Size
6x13.5x12inches
152.4 x 342.9 x 304.8mm
(HxWxD)
Weight <17lbs.(7.7kg)
Electrical
Power Supply
Input: Universal 100-240 VAC, 50-
60 hz
Output: 9 VDC
Battery
Sealed Lead Acid
6 VDC, 7.2 AH
General
Display
LCD Graphical 240 x 64
Pixels, Backlight
Ventilation Variable speed internal fan
Oscilloscope Output
Isolated (uncalibrated), BNC
Connector
Setup Memory EEPROM, All Parameters
Memory Retention 10 years w/o Power
Operating Range 15 to 35 °C
Storage Range -20to60°C
Humidity Limit 20 to 90 %, Non-Condensing
Connections
Communications USB, DB9
Loads: 4mm safety jacks
22
22
Comentários a estes Manuais