15
MAINS VOLTAGE
RANGE
90 to 264 VAC, 50/60 Hz
ACCURACY
± 3% of Reading, ± 1 digit
DEVICE CURRENT
RANGE
0 to 19.99 Amps, RMS
ACCURACY
± 5% of Reading, ± 1 digit
EARTH RESISTANCE
RANGE
0 to 19.99 Ω
ACCURACY
0 to 1.99 Ω
± 1% Range
2.00 to 19.99 Ω
± 1% Range
CURRENT SOURCE
10 mA
(Note: 1 Amp Current Source Available as an optional
accessory, see description section for ordering
information on the CS-2000 Series)
LEAKAGE CURRENT
RANGE
0 to 1999 µA, RMS
ACCURACY
DC
± 1% Reading, ±1 digit
25 Hz up to 1 kHz
± 1% Reading, ±1 digit
1 kHz up to 100 kHz
± 2.5% Reading, ±1 digit
100 kHz to 1 MHz
± 5% Reading, ± 1 digit
LEAKAGE LOAD
1000 Ω
AAMI ES1-1993 or IEC 601
User Selectable
FUSE
250 mA, 250 V, 5x20 mm, Fast Acting
(Receptacle Ground Leg)
BC Part No. UF-0250-01