case characters in the command summary table below are used to show the shortened
versions of the commands if available.
SA-2001-R Communication Command Summary
Keyword
Node
Value
Description
CONFigure
MODE
MVOLtage
DCURrent
ERESistance
EGROund
ENCLosure
EXTernal
Mains Voltage
Device Current
Earth Resistance
Earth / Ground Leakage
Enclosure Leakage
External Measurement
HOT
OPEN
CLOSed
Receptacle Hot Lead Open
Receptacle Hot Lead Closed
NEUtral
OPEN
CLOSed
Receptacle Neutral Lead Open
Receptacle Neutral Lead Closed
GROUnd
OPEN
CLOSed
Receptacle Ground Lead Open
Receptacle Ground Lead Closed
POLarity
FWD
REV
Receptacle Polarity Forward
Receptacle Polarity Reversed (reverse
HOT/NEUTRAL)
FILter
AAMI
IEC
Set AAMI test load
Set IEC test load
SYSTem
UNITs?
Returns the units of Measure (Volts, Amps, uA, Ohms)
[Read Only]
MEASurement?
Returns the latest measurement [Read Only]
KEY
MUP
Changes Mode Up one position
MDN
Changes Mode Down one position
HOT
Duplicates the HOT key press
NEUtral
Duplicates the Neutral Key press
GROUnd
Duplicates the Ground key press
POLarity
Duplicates the Polarity key press
MODel?
Returns Safety Analyzer Model [Read Only]
VERsion?
Returns Firmware Version [Read Only]