LOAD TABLE – Up to twelve Load Resistance Values (each with a range from 0.0 to
6,500.0 ohms) may be set in the FACTORY SETUP Configuration. These values are used
in combination with the Custom Load Sets to determine the Load Configuration Table.
These options are available if the Load Selection parameter is set to “Table”. The settings
will be the individual calibrated loads followed by the Load Sets. Since these values can
be set to the actual values of the real resistors, this allows for maximum accuracy in the
wattage calculations. By default, the following values are loaded into this table when the
ESU-2050 instruments ships from the factory:
Load #1: 10
Load #2: 20
Load #3: 30
Load #4: 50
Load #5: 100
Load #6: 200
Load #7: 300
Load #8: 500
Load #9: 1000
Load #10: 2000
Load #11: 3000
Load #12: 4000