BC Biomedical IPA-2000S Manual do Utilizador Página 58

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 162
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 57
2-26
PRIMING
PRIMINGPRIMING
PRIMINGPRIMING
PRIMING
Close Valve and Remove Priming Syringe
Once the 50 mL of Distilled Water is injected into the Module,
close the Valve and remove the syringe.
STEP 5
9
Press Start Key to Set the Prime
As prompted, use to set the prime and go to the test
selection screen.
STEP 6
9
CHANNEL 1 IDLE
SELECT TEST TYPE
FLOW TEST
Before selecting test, prepare and attach the DUT following
steps 7 through 9 to insure proper connection.
NOTE:
CFor Maximum Accuracy, all air
must be cleared from the unit. This
is
best done by a high flow rate which
is easily obtained with firm pressure
on the priming syringe.
STEP 4
9
Open Valve, Forcefully Inject Distilled Water
Open Valve and forcefully inject 50 mL of Distilled Water. If
more than one syringe full of water is required, close valve
between fills. Take care not to inject air into the unit.
Vista de página 57
1 2 ... 53 54 55 56 57 58 59 60 61 62 63 ... 161 162

Comentários a estes Manuais

Sem comentários