8
A 1 gram weight is supplied to check the calibration and programming. With the DUT
turned off, zero the unit. Place the weight on the arm of the cone target. Within 3
seconds the unit should read 14.6 ±0.2 Watts or 1.00 ±0.01 grams. If this reading is not
accurate, the USP-100A needs to be calibrated by BC Group.
It is recommended that the USP-100A be returned to BC Group on a routine yearly
basis for calibration and certification.