CAUTION - SERVICE
The SA-2010 is intended to be serviced only by
authorized service personnel. Troubleshooting
and service procedures should only be
performed by qualified technical personnel.
CAUTION - ENVIRONMENT
Exposure to environmental conditions outside
the specifications can adversely affect the
performance of the SA-2010. Allow SA-2010 to
acclimate to specified conditions for at least 30
minutes before attempting to operate it.
CAUTION - INSPECTION
The SA-2010 should be inspected before each
use for obvious signs of abuse or wear. The SA-
2010 should not be used and should be serviced
if any parts are in question.
CAUTION - USAGE
The SA-2010 is not a continuous duty device, it
is intended for short duration testing within the
current limits and duty periods specified. Do not
leave the DUT connected to the SA-2010 for
extended time periods.
Do not to drop the SA-2010.
WARNING - VOLTAGE
When the SA-2010 is in Lead Isolation mode and
the ISO key is depressed, 110% of line voltage is
applied to the Patient lead connectors and/or
External test leads. Although this voltage is
applied through an internal current limiting
resistance of 121 kΩ (per standard test
specifications), Do not touch the test leads,
connections or DUT while the ISO key is
depressed.